Laboratory autoclaves with heating
For pressurised steam sterilisation. Temperature control is monitored via a built-in PT 100 sensor. With 1 min. to 100 hours timer.
An accessory media probe (Cat. No. 9.842 070), also Pt 100, can be connected. If the “media probe” button is switched on, the autoclave cools down to 60°C after sterilisation and maintains this temperature until the unit is switched off. Culture media can therefore be processed more easily.
Chamber and lid made of stainless steel.
La-VA-MCS.
Fully automatic, 4 bar. With solenoid valve for automatically venting and pulsing steam when sterilising liquids. Safety catch is only temperature dependent when used with accessory media probe.
- Usable volume: 20L
- Chamber (dia. x H): 260mm x 400mm
- Supply requirements: 230V 50Hz, 1500W
Toote variatsioonid
Type | Temperature °C | External dimensions mm | Pressure bar | Catalog number | PK | Kogus | |
---|---|---|---|---|---|---|---|
LaM-3-20-MCS-J | 143 | 410 x 530 | 3 | 6.700 608 | 1.00 | ||
La-MCS-203 | 143 | 390 x 530 x 360 | 3 | 9.842 076 | 1.00 | ||
LaM-3-20-ECZ-J* | 143 | 410 x 595 | 3 | 9.842 066 | 1.00 | ||
La-MCS-204 | 151 | 390 x 530 x 360 | 4 | 9.842 077 | 1.00 | ||
La-VA-MCS | 151 | 390 x 530 x 360 | 4 | 9.842 079 | 1.00 |