Automatic Titrators TitroLine® 5000
Very compact titrator for simple routine titrations. Thanks to the standard methods the creating of methods is no longer required. Just load the method and press the start button. GLP compliant results with titration curve can be documented on a connected printer or USB-memory stick.
- Following titration modes are possible:
- Titrations to pH, mV and µA endpoint (2 EP)
- Titrations to one equivalence point (pH und mV) with dynamic or linear titration addition
- Manual titrations with the controller
- Dosing tasks
- Number of methods: 5
Typical applications for the TitroLine® 5000 are:
- Salt content in foodstuffs (cheese, soy sauce, ketchup)
- Total acidity in wine and beverages
- Nitrogen according to Kjeldahl
- Chloride in drinking water and other samples
- Alkalinity (p + m-value)
- FOS/TAC
Supplied with:
TL 5000/20 M1: Basic unit without electrode, with ready to use assembled 20 ml dosing unit TZ 3130 with hoses, fittings and titration tip, controller TZ 3880, titration clamp Z 305, stand rod TZ 1748 magnetic stirrer TM 50 and power supply 100-240 V
TL 5000/50 M1: as TL 5000/20 M1, only with ready to use assembled 50 ml dosing unit TZ 3160
TL 5000/20 M2: as TL 5000/20 M1, additionally with pH combination electrode and buffer set
TL 5000/50 M2: as TL 5000/20 M2, only with ready to use assembled 50 ml dosing unit TZ 3160
TL 5000/20 M3: as TL 5000/20 M1 additionally with Ag electrode and electrode cable
- Specifications
- Measuring input electrode pH/mV-input with DIN or BNC socket
- Measuring input Pt 1000/NTC 30: Temperature sensor connection for resistance probes Pt 1000 or NTC 30 (socket: 2 x 4 mm)
-
Interfaces: 1 x USB A und 1 x USB-B
2 x RS-232-C - Stirrer connection: TM 50 power supply directly through piston burette
- Keyboard: The unit is operated using the keys on the device itself, the controller TZ 3880 and optional PC-keyboard (USB)
- Display: graphics-capable TFT display
- Dosing units: 20 ml or 50 ml dosing unit, interchangeable
- Burette resolution: 8000
- Dosing accuracy: systematic error 0.15 %, random error 0.05 % in compliance with EN ISO 8655-6
- Supply requirements: 230V, 50/60Hz
- Conformity: ISO 8655, part 6
- CE-mark: EMC: 2004/108/EG; safety EG- Directive 2006/95
Toote variatsioonid
Type | Catalog number | PK | Kogus | |
---|---|---|---|---|
TL 5000/50 M1 | 9.863 185 | 1.00 | ||
TL 5000/50 M2 | 9.863 187 | 1.00 | ||
TL 5000/20 M2 | 9.863 186 | 1.00 | ||
TL 5000/20 M3 | 9.863 191 | 1.00 | ||
TL 5000/20 M1 | 9.863 184 | 1.00 |