Climate Chambers ICH C, with humidity and CO2 control

The climate chamber generation with the unique ControlCOCKPIT for a quick and intuitive operation. Unparalleled temperature and humidity homogeneity for long-term stable ambient conditions.

  • PID microprocessor control with integrated auto-diagnostic system
  • Interior stainless steel 1.4301 (ASTM 304), deep-drawn
  • Structured stainless steel housing, rear of zinc-plated steel
  • High-Operation and adjustment of all parameters directly at the ControlCOCKPIT temperature connectors on the rear of the appliance for single-phase power connection according to country specific systems and IEC standards
  • Internal data logger with a storage capacity of at least 10 years
  • German, English, French, Spanish language settings available on the ControlCOCKPIT
  • Digital backwards counter with target time setting adjustable between 1 minute and 99 days
  • The SetpointWAIT function guarantees that the process time does not start until the set temperature is reached at all measuring points – optional for temperature values recorded by the freely positionable Pt100 sensors inside the chamber
  • Adjustment of three calibration values for temperature and additional appliance specific parameters directly at the ControlCOCKPIT, for example relative humidity
  • Moisture tight standard entry port, at the back, 40 mm inner diameter, can be closed by silicone stopper

TwinDISPLAY: ControlCOCKPIT with two TFT colour displays

  • Available parameters on the ControlCOCKPIT: Temperature (Celsius or Fahrenheit), programme time, relative humidity, time zones, summertime/wintertime
  • Two Pt100 sensors DIN class A in a 4-wire circuit for mutual monitoring, taking over functions in case of an error
  • AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
  • ControlCOCKPIT with USB port for uploading programmes, reading out protocol logs, activating the User-ID function
  • Displaying of already logged protocol data on the ControlCOCKPIT (max 10,000 values correspond to approx. 1 week)
  • Ethernet interface on the rear of the appliance for reading out the protocol log and for uploading programmes and for online logging
  • Multiple overtemperature protection: Electronic temperature monitoring TWW/TWB (protection class 3.3 or 2 and mechanical temperature limiter TB (protection class 1) acc. to DIN 12880, AutoSAFETY automatically adjusts to the set value within a freely adjustable tolerance range. Setting individual MIN/MAX values for over/undertemperature alarm

Toote variatsioonid

CO2 concentrationDepth int.External dimensions (WxDxH)Ext. depthExt. heightExt. widthHeight int.Internal dimensions (WxDxH)Int. widthMaterial insertsMaterial interiorMax. CO₂Max. no. of insertsMax. temperatureMax. voltageMin. CO₂Min. temperatureNominal capacityCatalog numberPower supplyRated capacityTemperatureTypeWeightWith CO2 controlWith timerPKKogus 
0 ... 20 %500 mm824 x 685 x 1,552 mm685 mm1552 mm824 mm800 mm640 x 500 x 800 mm640 mmStainless steelStainless steel20 %960 °C230 V1277630 °C256 l6.261 039230 V, 50...60 Hz1350 W0 ... 60 °CICH260C160 kgYesYes1.00
0 ... 20 %600 mm1,224 x 785 x 1,950 mm785 mm1950 mm1224 mm1200 mm1,040 x 600 x 1200 mm1040 mmStainless steelStainless steel20 %1460 °C230 V1277630 °C749 l6.261 060230 V, 50...60 Hz1350 W0 ... 60 °CICH750C249 kgYesYes1.00
0 ... 20 %400 mm745 x 585 x 1,233 mm585 mm1233 mm745 mm480 mm560 x 400 x 480 mm560 mmStainless steelStainless steel20 %560 °C230 V1277630 °C108 l6.261 037230 V, 50...60 Hz1350 W0 ... 60 °CICH110C114 kgYesYes1.00