Climate Chambers ICH, with humidity control

The climate chamber generation with the unique ControlCOCKPIT for a quick and intuitive operation. Unparalleled temperature and humidity homogeneity for long-term stable ambient conditions.

  • PID microprocessor control with integrated auto-diagnostic system
  • Interior stainless steel 1.4301 (ASTM 304), deep-drawn
  • Structured stainless steel housing, rear of zinc-plated steel
  • High-Operation and adjustment of all parameters directly at the ControlCOCKPIT temperature connectors on the rear of the appliance for single-phase power connection according to country specific systems and IEC standards
  • Internal data logger with a storage capacity of at least 10 years
  • German, English, French, Spanish language settings available on the ControlCOCKPIT
  • Digital backwards counter with target time setting adjustable between 1 minute and 99 days
  • The SetpointWAIT function guarantees that the process time does not start until the set temperature is reached at all measuring points – optional for temperature values recorded by the freely positionable Pt100 sensors inside the chamber
  • Adjustment of three calibration values for temperature and additional appliance specific parameters directly at the ControlCOCKPIT, for example relative humidity
  • Moisture tight standard entry port, at the back, 40 mm inner diameter, can be closed by silicone stopper

TwinDISPLAY: ControlCOCKPIT with two TFT colour displays

  • Available parameters on the ControlCOCKPIT: Temperature (Celsius or Fahrenheit), programme time, relative humidity, time zones, summertime/wintertime
  • Two Pt100 sensors DIN class A in a 4-wire circuit for mutual monitoring, taking over functions in case of an error
  • AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
  • ControlCOCKPIT with USB port for uploading programmes, reading out protocol logs, activating the User-ID function
  • Displaying of already logged protocol data on the ControlCOCKPIT (max 10,000 values correspond to approx. 1 week)
  • Ethernet interface on the rear of the appliance for reading out the protocol log and for uploading programmes and for online logging
  • Multiple overtemperature protection: Electronic temperature monitoring TWW/TWB (protection class 3.3 or 2 and mechanical temperature limiter TB (protection class 1) acc. to DIN 12880, AutoSAFETY automatically adjusts to the set value within a freely adjustable tolerance range. Setting individual MIN/MAX values for over/undertemperature alarm

Toote variatsioonid

Depth int.External dimensions (WxDxH)Ext. depthExt. heightExt. widthHeight int.Internal dimensions (WxDxH)Int. widthMaterial insertsMaterial interiorMax. no. of insertsMax. temperatureMax. voltageMin. temperatureNominal capacityCatalog numberPower supplyRated capacityTemperatureTypeWeightWith CO2 controlWith timerPKKogus 
600 mm1,224 x 785 x 1,950 mm785 mm1950 mm1224 mm1200 mm1,040 x 600 x 1200 mm1040 mmStainless steelStainless steel1460 °C230 V-10 °C749 l9.538 020230 V, 50...60 Hz1350 W-10 ... 60 °CICH750249 kgNoYes1.00
400 mm745 x 585 x 1,233 mm585 mm1233 mm745 mm480 mm560 x 400 x 480 mm560 mmStainless steelStainless steel560 °C230 V-10 °C108 l9.538 018230 V, 50...60 Hz1350 W-10 ... 60 °CICH110109 kgNoYes1.00
500 mm824 x 685 x 1,552 mm685 mm1552 mm824 mm800 mm640 x 500 x 800 mm640 mmStainless steelStainless steel960 °C230 V-10 °C256 l9.538 019230 V, 50...60 Hz1350 W-10 ... 60 °CICH260160 kgNoYes1.00