Polarimeter SAC-i 589/882
Measurement Scales: Angle of rotation, specific rotation, International Sugar Scale, concentration, purity.
- Wavelength measurement at 589 nm (equivalent to the sodium-D line spectrum) and 882 nm (near infra red)
- Measurement stability in 13 seconds (4 seconds in the high-speed and consecutive measurement mode)
- Resolution of up to 3 decimal places
- Repeatability of ±0.003° Angle of Rotation and ±0.009°Z
- Up to 999 automatic consecutive measurements followed by average value display
- Sleep and timer feature to take measurement later at specified time
- Improved usability due to touch screen technology
- Connectivity to RX Digital Refractometers for automatic purity measurement
- Specifications
-
Measurement range
Angle of rotation:
International standard sugar scale: -89.999 to +90.000° or -360 to +360°
-259 to +259° -
Resolution
Angle of rotation:
International standard sugar scale:
0.001°
0.001°Z -
Measurement accuracy
Angle of rotation 589 / 882 nm:
International standard sugar scale 589 / 882 nm:
±0.005° / ±0.010° (-45 to +45°)
±0.015°Z / ±0.0700°Z (-130 to +130°Z) - Measurement temperature range: 5 to 40°C
- Observation tubes: 100mm (5ml), 200mm (10ml)
- Power supply: 100-240V, 50/60Hz a.c.
-
Dimensions:
Weight: 600 x 365 x 210mm
20kg
Toote variatsioonid
Type | Catalog number | PK | Kogus | |
---|---|---|---|---|
SAC-i 589/882 | 6.286 467 | 1.00 |