Polarimeter SAC-i 589/882

Measurement Scales: Angle of rotation, specific rotation, International Sugar Scale, concentration, purity.

  • Wavelength measurement at 589 nm (equivalent to the sodium-D line spectrum) and 882 nm (near infra red)
  • Measurement stability in 13 seconds (4 seconds in the high-speed and consecutive measurement mode)
  • Resolution of up to 3 decimal places
  • Repeatability of ±0.003° Angle of Rotation and ±0.009°Z
  • Up to 999 automatic consecutive measurements followed by average value display
  • Sleep and timer feature to take measurement later at specified time
  • Improved usability due to touch screen technology
  • Connectivity to RX Digital Refractometers for automatic purity measurement
  • Specifications
  • Measurement range
    Angle of rotation:
    International standard sugar scale:
    -89.999 to +90.000° or -360 to +360°
    -259 to +259°
  • Resolution
    Angle of rotation:
    International standard sugar scale:

    0.001°
    0.001°Z
  • Measurement accuracy
    Angle of rotation 589 / 882 nm:
    International standard sugar scale 589 / 882 nm:

    ±0.005° / ±0.010° (-45 to +45°)
    ±0.015°Z / ±0.0700°Z (-130 to +130°Z)
  • Measurement temperature range: 5 to 40°C
  • Observation tubes: 100mm (5ml), 200mm (10ml)
  • Power supply: 100-240V, 50/60Hz a.c.
  • Dimensions:
    Weight:
    600 x 365 x 210mm
    20kg

Toote variatsioonid

TypeCatalog numberPKKogus 
SAC-i 589/8826.286 4671.00