Steriliser SFplus

Safer than disinfection, hot-air sterilisation of containers and instruments.
The safety feature for this hot air steriliser: Setpoint Wait. This means that the programme for sterilisation only starts when the set temperature has been reached. This feature can also be used with freely positionable Pt100 temperature sensors. Here the sterilisation time only begins when the set temperature has been reached at all measurement points, and reliable sterilisation is guaranteed at all times.

Textured stainless steel housing, rear of zinc-plated steel, intuitively operated TwinDISPLAY(TFT colour display) with touchscreen
You can find a detailed description of the “TwinDISPLAY” model at www.memmert.com

  • Available parameters on the ControlCOCKPIT: Temperature (Celsius or Fahrenheit), programme time, relative humidity, time zones, summertime/wintertime
  • Two Pt100 sensors DIN class A in a 4-wire circuit for mutual monitoring, taking over functions in case of an error
  • AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
  • ControlCOCKPIT with USB port for uploading programmes, reading out protocol logs, activating the User-ID function
  • Displaying of already logged protocol data on the ControlCOCKPIT (max 10,000 values correspond to approx. 1 week)
  • Ethernet interface on the rear of the appliance for reading out the protocol log and for uploading programmes and for online logging
  • Multiple overtemperature protection: Electronic temperature monitoring TWW/TWB and mechanical temperature limiter TB acc. to DIN 12880, AutoSAFETY automatically adjusts to the set value within a freely adjustable tolerance range. Setting individual MIN / MAX values for over/undertemperature alarm

Toote variatsioonid

With timerCatalog numberPKTemperatureDepth int.Ext. depthExt. heightExt. widthHeight int.Int. widthMax. no. of insertsMax. temperatureMax. voltageMin. temperatureNominal capacityPlug typeRated capacityTypeWeightKogus 
Yes9.842 5321.00RT330 mm514 mm944 mm585 mm560 mm400 mm6250 °C230 V-10 °C74 lEU2.5 kWSF75plus66 kg
Yes9.842 5301.00RT250 mm434 mm704 mm585 mm320 mm400 mm3250 °C230 V-10 °C32 lEU1.6 kWSF30plus46 kg
Yes9.842 5311.00RT330 mm514 mm787 mm585 mm400 mm400 mm4250 °C230 V-10 °C53 lEU2.0 kWSF55plus57 kg
Yes9.842 5331.00RT400 mm584 mm864 mm745 mm480 mm560 mm5250 °C230 V-10 °C108 lEU2.8 kWSF110plus74 kg