Temperature/conductivity data logger EBI 12-TC230
Temperature/conductivity data logger for regular measurement of conductivity during the cleaning and disinfection process in the washer-disinfector. The conductivity of the final rinse water is an indicator of the quality of the cleaning performance.
- Data logger can be sterilised
- 1 external PT1000 temperature sensor
- 1 external conductivity sensor
- Conductivity measurement range: 1 to 2000 µS/cm
- Conductivity resolution: 0.1 µS/cm
- Storage modes: Endless measurement, measurement upon start time, measure immediately until end of memory, start/stop measurement
- Memory for up to 50160 measured values per measurement channel
- Stainless steel housing
- Factory calibration certificate: 0 °C, 60 °C, and 121 °C; 5 µS/cm, 100 µS/cm, and 1413 µS/cm
Toote variatsioonid
| Battery/rechargeable battery | Ø | Dimensions (ØxH) | Height | IP code | Material housing | Max. measuring cycle | Max. temperature | Min. measuring cycle | Min. temperature | Catalog number | Resolution | Temperature | Type | Weight | PK | Kogus | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|